Product name:LC600B Gradient Liquid Chromatograph -
Order number: -
Model:LC600B -
Stock status:Delivery according to contract -
Logistics:by sea, by air, by express, etc.
- Product Introduction
-
1. Product introduction
The LC-600B intelligent full-control liquid chromatography system consists of two P600C high-pressure constant-flow pumps, one UV600B ultraviolet detector and one SCL600B system controller to directly form a gradient analysis system. The infusion pump adopts the plunger floating installation method to enhance the stability of the instrument performance and the durability of use. The instrument is easy to operate and has a variety of configurations. It is suitable for application analysis in various industries and is convenient for customers to choose.
2. Product parameters
LC600B high pressure constant flow infusion pump
2. Maximum infusion pressure: 0~9999Psi
3. Flow setting range: 0.001~9.999mL/min (adjust the flow in steps of 0.001mL/min)
4. Flow setting value error: ≤0.5%
5. Flow stability error: ≤0.2%RSD
6. Gradient error: ±1% (0~100% water/acetone aqueous solution 2-liquid gradient)
7. Pressure pulsation: less than 15Psi (flow 1mL/min, pressure 600-1600Psi)
8. Pump sealing: the pressure is 5400Psi, the time is 10min, and the pressure drop is less than 400Psi
9. Time program function: yes
10. Size: W260×H130×D420mm
11. Weight: 11kg
12. Ambient temperature range: 4~40℃
UV-Vis Variable Wavelength Detector
14. Wavelength indication error ≤±1nm
15. Wavelength repeatability error ≤±0.1nm
16. Dynamic noise ≤ ±0.75×10-5AU (methanol, 1ml/min, 254nm, 20℃)
17. Static noise ≤ ±0.5×10-5 AU (empty cell, response time 1 second, 20℃)
18. Dynamic baseline drift ≤±1×10-4AU/h (methanol, 1mL/min, 254nm, 20℃)
19. Static baseline drift ≤ 5×10-4 AU/h (empty cell, response time 1 second, 20℃)
20. Linear range ≥104
21. Minimum detection concentration ≤1×10-9g/mL (naphthalene/methanol solution)
22. Qualitative repeatability RSD6≤ 0.1%
23. Quantitative repeatability RSD6≤ 0.5%
24. Spectral bandwidth 6nm
25. The volume of the flow cell is 8 μL
26. Optical path 10mm
27. Time program function Yes
28. Dimensions W260×H130×D420mm
29. Weight 11kg
30. The operating ambient temperature range is 4~40℃
3. Product Features
The hardware has the VP function, records maintenance information and operation records, and meets the requirements of GLP/GMP; the system controller has the functions of humanized design such as clock, thermometer, and hygrometer.
2. Superior performance - to meet the strict requirements of customers for the instrument
The infusion pump adopts the plunger made by considering the material structure, the reinforced Teflon sealing ring, and the floating installation method of the plunger, which enhances the stability of the instrument performance and the durability of use.
3. The detector adopts imported deuterium lamp, photocell and double-beam monochromator composed of 1200/mm concave grating, precision-machined double-lens flow cell, high-precision microprocessor that controls wavelength adjustment, and dual-channel high-speed sampling frequency to ensure It has the characteristics of low noise, low drift and high sensitivity.
4. The mixer mixes evenly, and the baseline collected using the gradient program maintains good reproducibility.
5. Reliable results—meet the accurate requirements of customers for results, and do comparative tests with imported instruments, and the analysis results are highly consistent.
6. Simple operation - it is convenient for customers to operate the software proficiently. The software adopts multi-window mode, which is easy to operate.
7. Exquisite appearance - to meet the visual requirements of customers for the instrument, the appearance is exquisite, bringing visual enjoyment.
8. a variety of user software - to meet the needs of different users
The chromatographic data workstation dedicated to the LC600B isocratic liquid chromatograph is embedded in the detector. There are three sets of VI5890, N2000, and N3000 for customers to use selectively to meet the needs of different users.