/uploadfiles/site/img/20220304/6221a52bdbbcd.png

PRODUCTS CENTER

Service Hotline
+86-532-80921517
Atmos./parti.sampler
Atmosphere/TSP Sampler
Atmosphere/TSP Sampler
Atmosphere/TSP Sampler
Atmosphere/TSP Sampler
JCH-6120-1 Atmosphere/TSP Comprehensive Sampler
  • Product name:JCH-6120-1 Atmosphere/TSP Comprehensive Sampler
  • Order number:
  • Model:JCH-6120-1
  • Stock status:Delivery according to contract
  • Logistics:by sea, by air, by express, etc.
  • Product Introduction
  • 1. Product introduction


    JCH-6120 intelligent atmospheric /particulate matter comprehensive sampler (hereinafter referred to as the sampler) is used to collect total suspended particles (TSP, PM10 , PM2.5, etc.) and various gas components (SO2, NOx , etc.) in the atmosphere etc.) sample sampler. The instrument uses the solution absorption method to collect various harmful gases in the ambient atmosphere and indoor air; uses the filter membrane weighing method to capture the total suspended particulates (TSP) and inhalable particles (PM10) or (PM2.5) in the ambient atmosphere. . The new upgrade of the three core systems ensures the reliability of the instrument, improves the stability of the system, and enhances the accuracy of control. There are great improvements in small portability and flow stability, which greatly reduces labor intensity. It can be used for routine and emergency monitoring of gaseous substances and aerosols in environmental protection, health, labor, safety supervision, military, scientific research, education , testing companies and other departments.


    2. Product parameters


    The main parameters

    parameter range

    Resolution

    Accuracy

    Atmospheric sampling flow

    ( 0.1 ~ 1.0 ) L/min

    Can be customized 0.1-2L/min

    0.1L/min

    better than ± 2 .5 %

    Particulate sampling flow

    ( 60 ~ 130 ) L/min

    0.1L/min

    better than ± 2 .5 %

    delay time

    1min ~99h59min

    1min

    Better than ± 0.2%

    sampling time

    1min ~99h59min

    1min

    Better than ± 0.2%

    Intervals

    1min ~99h59min

    1min

    Better than ± 0.2%

    Equally spaced sampling times

    1 to 99 times

    Intervals

    <99h59min

    MAX sample volume

    9999.99L

    pre-gauge pressure

    ( -20 ~0 ) Kpa

    0.01Kpa

    better than ± 2.5%

    atmospheric pressure

    ( 70 ~ 130 ) Kpa

    0.1Kpa

    better than ± 2.5%

    Temperature control range

    ( 15 ~30 )℃

    0.1 ℃

    better than ± 1.5 ℃

    Operating temperature

    (-30 ~+50) ℃

    0.1 ℃

    better than ± 1 ℃

    Working power

    AC220V ± 10% 50Hz

    PM2.5 cutting characteristics

    Da50 = (2.5 ± 0.2) μm σ g = ( 1.2 ± 0.1 ) μm

    PM10 cutting characteristics

    Da50 = (10 ± 0.5) μm σ g = ( 1.5 ± 0.1 ) μm

    Inlet velocity

    0.3m/s

    b/a

    0.625

    Sampling flow

    100L/min

    Effective membrane diameter

    Ф 80mm

    Connector

    M20 × 1.5

    Dimensions

    400mm × 270mm × 200mm

    Total Weight

    about 5 kg

     

    3. Product Features


    1. Electronic flowmeter automatically and accurately controls flow, no flow fluctuation, constant current sampling, excellent flow calibration software

    2. The atmosphere has a constant temperature automatic heating function

    3. The atmosphere uses high-performance ultra-bass imported diaphragm pump, which greatly improves stability, long service life and ultra-low noise

    4. The high-efficiency anti-suction desiccator design for atmospheric use can effectively prevent the absorption liquid from being sucked back due to misoperation, and enhance the safety of the instrument

    5. The atmosphere uses high-quality dust filter elements,Realize the function of particulate matter filtering, prevent entering the gas path to interfere with sampling, and protect the sampling pump and gas path at the same time

    6. Atmospheric sampling two-way design, flexible sampling method, can collect two kinds of gases or parallel samples, AB way is controlled separately

    7. Atmospheric sampling adopts red and blue dual-color gas connection pipes, which can be easily and accurately connected to the gas path

    8. Particle sampling adopts brushless control fan, with low noise and large load, suitable for continuous long-term work
    9. TSP/PM10/PM5/PM2.5 (optional) sampling head is made of aluminum alloy material, anti-static adsorption

    10. Real-time monitoring of pressure and temperature, automatic compensation of flow deviation, and further optimization of flow accuracy

    11. The Chinese LCD screen with automatic contrast adjustment is suitable for sampling in cold areas, and the popular software display interface realizes good human-computer interaction
    . 12. Automatically calculates the cumulative sampling volume, and converts the standard sampling volume according to air pressure and temperature at the same time.
    13. Atmospheric pressure can be Input and measure for use in low pressure environments

    14. Optional JCD-1500 upgraded mobile power box can realize normal sampling without AC power

    15. A set of sample storage device is optional, which is convenient for sample collection and storage.