PRODUCTS CENTER
-
Product name:JC-YQ-17 Pollution source exhaust gas multifunctional sampling tube -
Order number: -
Model:JC-YQ-17 -
Stock status:Delivery according to contract -
Logistics:by sea, by air, by express, etc.
- Product Introduction
-
1. Technical features
Integrated whole process heating bong with beautiful appearance;
sampling tube is lined with PTFE tube to avoid pipeline adsorption;
filter membrane holder is made of PTFE material, which automatically maintains a constant temperature of 120 ℃ during sampling;
supporting tube Circuit closing device, which is convenient for closing the gas circuit;
high-efficiency incubator, which can be placed in various combinations of absorption bottles;
can be used for both horizontal flue sampling and vertical flue sampling;
low-voltage DC power supply, safe and reliable use;
imported high-precision temperature Controller, high-power power supply, fast heating to constant temperature;
multi-function combined sampling gun, including smoke temperature and pitot tube flow rate measurement function;
multi-purpose gun, can collect hydrochloric acid, sulfuric acid, hydrogen fluoride (fluoride), bromide Hydrogen, chromate and other pollutants;
2. Technical parameters
The main parameters
parameter
constant temperature
(80 ~ 200)℃, error ±3℃
Heating time
<20min
Sampling tube length
1.5m
Sampling nozzle diameter
Φ4.5/Φ6/Φ7/Φ8/Φ10/Φ12
membrane diameter
Φ47
Absorption bottle specifications
250ml/75ml impact absorption bottle
power supply
DC24V
Power consumption
<400W
weight
≈8.0kg