Product name:JC-GGC400 water quality sulfide-acidification blowing instrument -
Order number: -
Model:JC-GGC400 -
Stock status:Delivery according to contract -
Logistics:by sea, by air, by express, etc.
- Product Introduction
-
1. Product introduction
1. Product principle
The sulfide in the water sample is acidified, and the generated hydrogen sulfide enters the absorption bottle/absorption color developing tube with the carrier gas (nitrogen) and is absorbed by the absorption solution (zinc acetate-sodium acetate solution or 2% sodium hydroxide solution). The analytical method is used to analyze and measure the sulfide ions absorbed in the absorption bottle/absorption color developing tube.
2. Product use
The JC-GGC400 water quality sulfide-acidification gas blowing instrument is developed and produced according to the standard of water quality sulfide determination-methylene blue spectrophotometry (iodometric method). Meet the sample pretreatment needs of water sulfide determination.
3. Product implementation standards
HJ 1226—2021 "Determination of Sulfur in Water Quality Methylene Blue Spectrophotometry"
GB/T16489-1996 "Determination of Sulfide in Water Quality--Methylene Blue Spectrophotometry"
HJ/T600-2000 "Determination of Sulfide in Water Quality - Iodometric Method"
State Environmental Protection Administration compiled "Water and Wastewater Monitoring and Analysis Methods (Fourth Edition)"
2. Product parameters
1. Number of samples: 4 or 6 (standard 4, 6 customized)
2. Heating method: automatic temperature control constant temperature water bath
3. Heating power: l000W, even heating
4. Nitrogen flow meter control: 0-10L/min
5. Temperature range: room temperature -99.9℃
6. Display mode: digital display
7. Temperature control accuracy: ±1℃
8. Nitrogen inlet pressure: 0.1Mpa
9. Nitrogen flow branch: 0.06-0.6L/min
10. Working power supply: AC(220±22)V, 50HZ
11. Ambient temperature: (5-35) ℃
12. Ambient humidity: (0-95)%RH
3. Product Features
1. The large-capacity reaction bottle can process multiple samples at the same time.
2. The hydrochloric acid port, gas inlet and outlet are integrated and independent of each other, and the operation will not interfere .
3. Constant temperature water bath heating, with fast automatic constant temperature control system, the heating temperature can be manually set.
4. Adjust the sample rack and the lifting and rotating sample rack of the gas distribution system by the lifting and lowering position fixed knob. The sample is installed on the front, and the operation is convenient.
5. The nitrogen flow of each sample is independently controlled, adjusted or closed, and the needle valve gas flowmeter can accurately control and display the total gas consumption.