PRODUCTS CENTER
-
Product name:JC-YZ-600 Flame Graphite Furnace Integrated Atomic Absorption Spectrophotometer -
Order number: -
Model:JC-YZ-600 -
Stock status:Delivery according to contract -
Logistics:by sea, by air, by express, etc.
- Product Introduction
- Details Page
-
1.Product introduction
1. Optical system
wavelength range
185nm—910nm automatic adjustment
Grating line density
1800 lines/mm
Monochromator
Aberration Czerny-Turner Monochromator
Spectral bandwidth
0.1, 0.2, 0.4, 1.0, 2.0nm, multi-speed automatic switching
Wavelength accuracy
±0.10nm
wavelength repeatability
≤0.05nm
baseline drift
Static ±0.002A/30min Dynamic ±0.006 A/30min
Detector
Wide spectral range photomultiplier tube
Resolution
To distinguish 279.5nm and 279.8nm double lines of manganese lamps , the valley energy between the two spectral lines should be less than 20%
2. Light source system
lamp holder
8 lamp holder automatic switching
Lamp power supply mode
400Hz square wave pulse
Lamp current adjustment range
2 — 10mA average current
3. Atomic system
flame analysis method
Detection limit (Cu)
<0.003μg/ml
Characteristic concentration (Cu)
<0.015μg/ml/1%
burner
100mm full titanium burner
Precision
RSD<0.5%
sprayer
Metal Sleeve High Efficiency Glass Sprayer
spray chamber
Corrosion-resistant all-plastic spray chamber
safety measures
With a variety of automatic safety protection functions: acetylene leakage alarm, automatic shutdown system, etc.
Flameless (graphite furnace) analytical method
Detection limit ( Cd)
<0.4*10ˉ12 g
Feature quantity ( Cd)
<0.5*10ˉ12 g
Hydride Generation - Atomic Absorption
4. Flame analysis method
Measurement methods
Air -acetylene flame method
Concentration calculation method
Standard curve method, standard addition method, interpolation method.
Repeat measurement times
From 1 to 99 times, the average value, content, standard deviation and relative standard deviation are automatically calculated.
result print
Print test data or final analysis report. The results can be exported to a word worksheet.
Software Environment
Windows XP/ Windows 7 operating system, Chinese professional software
5. Flameless (graphite furnace) analysis method
Measurement methods
Graphite furnace peak height method, graphite furnace peak area method
Concentration calculation method
Standard curve method, standard addition method, interpolation method.
background correction
Deuterium lamp background correction is greater than 60 times, and self-absorption deduction background is greater than 60 times
Repeat measurement times
1-30 times, calculate the mean, give the standard deviation and relative standard deviation
result print
Print test data or final analysis report. The results can be exported to a word worksheet.
Software Environment
Windows XP/ Windows 7 operating system, Chinese professional software
2. Product parameters
power supply
AC 220V/50HZ
power
150VA
Instrument net weight
55kg
Instrument gross weight
80kg
Instrument volume
1110mm (horizontal) × 400mm (depth) × 500mm (height)
package volume
1350mm (horizontal) × 600mm (depth) × 850mm (height)
Working temperature
10~35℃
Working environment humidity
≤80%
3. Product Features
1. Total reflection optical system, full-band achromatic, automatic wavelength scanning and automatic peak finding.
2. With deuterium lamp button background, self-priming button background.
3. The uniquely designed all-titanium burner has smooth combustion and fast heat balance.
4. Automatic adjustment of negative high voltage, lamp current, automatic adjustment of the best element lamp position, automatic ignition, automatic flameout protection. 8 lamp stands (optional 2 high-performance lamps) automatic control system, which can adjust the hollow cathode lamp to the best position in all directions, can set the lamp position of a specific element independently, and can freely set the working lamp, preheating The lamp saves the element lamp preparation time to the greatest extent possible.
5. Graphite furnace heating: users can freely set the heating curve according to their needs. Advanced graphite furnace constant power temperature control design: the temperature of the graphite tube is uniform and consistent, and the working temperature of the graphite tube maintains a good dynamic stability to ensure the accuracy of the analysis.
-